Database, - Beckman Coulter LH 780 Instructions Manual

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

ADJUSTMENT PROCEDURES
ENABLING/DISABLING BLOOD DETECTOR
8.
A window appears with the message, "the system will be restarted to reinitialize with the
new settings".
9.
Select OK. The system shuts down windows. Another window appears with the message,
" It is now safe to turn off your computer."
10. Press the Power Off button on the computer tower.
11. Remove disk. Wait one minute and then Power On.
12. Log onto your system.
Archive Complete Database
1.
Format a CD for direct archiving.
2.
Select
3.
Select
4.
Ensure Archive Complete Database is selected, and then select
5.
Specify the location where you want to archive the database, and then select
4.6
ENABLING/DISABLING BLOOD DETECTOR
1.
Go to the Analyzer screen.
2.
Press
MAIN MENU
3.
Press
SYSTEM CONFIGURATION
4.
Press
BLOOD DETECTOR
Note: The system flags all parameters with a P (partial aspiration) when samples are
analyzed with blood detectors disabled. P flagged control run results are automatically
removed from statistical calculations, and there is no way to reintroduce the flagged runs
into the calculations.
5.
Press one of the following:
SYSTEM RUN
RETURN
4.7
SETTING UP ANALYZER DATE FORMAT
1.
Go to the Analyzer screen.
2.
Press
MAIN MENU
3.
Press
ANALYZER FUNCTIONS
4.
Press
DATE AND TIME
5.
Press
FORMAT
MM/DD/YY
r
DD/MM/YY
r
YY/MM/DD
r
where MM = Month, DD= Day, and YY= last two digits of the Year
4-4
on the Command Center to display the System Setup window.
to archive the database.
.
.
until the appropriate setting appears.
To cycle samples.
To display the Analyzer MAIN MENU screen.
.
.
.
until the date format you want to use appears:
.
.
PN 773023AC