TROUBLESHOOTING
DILUTER FUNCTIONS
F23 - Dispensing Diff Lytic Reagent Low Volume
1.
Go to the Numeric Keypad.
2.
Press
Numeric Keypad displays DIFF PAK LYSE LOW VOLUME while the instrument performs
this function. When the function completes processing, the Numeric Keypad displays
FUNCTION = 23. Press
3.
Press
F24 - Dispensing Diff Lytic Reagent High Volume
1.
Go to the Numeric Keypad.
2.
Press
Numeric Keypad displays DIFF PAK LYSE HIGH VOLUME while the instrument
performs this function. When the function completes processing, the Numeric Keypad
displays FUNCTION = 24. Press
3.
Press
F21 - Dispensing Diff Preservative Low Volume
1.
Go to the Numeric Keypad.
2.
Press
Numeric Keypad displays DIFF PAK PRESERVE LV DISPENSE while the instrument
performs this function. When the function completes processing, the Numeric Keypad
5-12
E
N
T
E
2
3
F
R
to dispense Diff lytic reagent into the mixing chamber. The
E
N
T
E
R
to perform this function again.
STOP
to exit this function. The Numeric Keypad displays READY.
E
N
T
E
F
2
4
R
to dispense Diff lytic reagent into the mixing chamber. The
STOP
to exit this function. The Numeric Keypad displays READY.
E
N
T
E
F
2
1
R
to dispense Diff preservative into the mixing chamber. The
E
N
T
E
R
to perform this function again.
PN 773023AC