CLEANING PROCEDURES
CLEANING BLOOD SAMPLING VALVE (REMOVE AND CLEAN INSIDE)
D Return BSV to its Original Location
IMPORTANT
section to the mounting post, moisten the inside surfaces with distilled water. Never use your fingers to
spread water over the surfaces because there could be oil or powder on your gloves.
KEYWAY
1
2-20
Incorrect results can be reported if oil or powder gets on the BSV. Before returning each
NOTCH
GUIDE
POST
2
TOP VIEW
1.
Replace the plastic washer on the
mounting post.
2.
Return the inner section to its original
position; align the notch with the
guide post and push the section into
place.
3.
Make sure to position all tubing
attached to the inner section behind
the probe.
4.
Ensure the routing of tubing is smooth
to prevent kinks.
5.
Align the key of the center section
with the keyway on the mounting post
and push it into place.
6.
Manually move the air cylinder block
as needed to position the key and
keyway.
7.
Ensure the routing of tubing is smooth
to prevent kinks.
CAUTION
Risk of system damage. The tubing
can be damaged if positioned in front of the
aspirator tip. Before mounting the front section,
make sure all tubing attached to the front
section is positioned behind the aspirator tip.
8.
Align the notch in the front section
with the guide post and push the
section into place.
9.
Ensure the routing of tubing is smooth
to prevent kinks.
PN 773023AC