5.14 CLOSING THE DILUTER LOWER COVER
1.
Lift the cover upward.
2.
Push the cover back toward the instrument.
5.15 AUTOSTOP
The LH Workstation software allows you to configure the system to automatically halt under
various conditions. By configuring the system to halt, it avoids analyzing samples during a
fault condition that could affect parameter results or slide preparation. This minimizes the
necessity to re-analyze those samples after the fault condition is resolved.
The number of runs (batches) configured for AutoStop needs to take into consideration that
the next sample has already been aspirated before data from the preceding sample has been
received at the Workstation. For example, AutoStop is configured to five and five consecutive
Partial Aspirations occur. The last six samples are potentially affected.
The LH 700 Series AutoStop is enabled within System Setup-[Quality Assurance], Control
Setup, for individual control lots. For controls, AutoStop must be enabled for the lot within
Control Setup and in Run Configuration, for control errors to trigger an AutoStop.
PN 773023AC
1
2
TROUBLESHOOTING
CLOSING THE DILUTER LOWER COVER
5
5-41