TROUBLESHOOTING
WORKSTATION MESSAGES
Table 6.22 Workstation Messages - N
Workstation
Log Message and Icon on Command Center
New order for Operator <Patient ID>, (<Sample ID>,
<Cass/Poss>) received from host
New patient information for <Patient ID> received from host
with order for [<Sample ID>, <Cass/Poss>], (<Last Name>,
<First Name>)
No Calibrator lot number. No statistics will be calculated
No latex Setup Information available, latex sample will not be
stored.
NO MATCH: Sample run received does not match any ToDo
list entry (<Run C/P>, <Run SID>).
Yellow traffic light is displayed until AutoStop condition is
reached.
Red traffic light is displayed when AutoStop condition is
reached.
NO READ: Sample run received without proper identification
(<Run C/P>, <Run SID>).
Yellow traffic light is displayed until AutoStop condition is
reached.
Red traffic light is displayed when AutoStop condition is
reached.
6-62
Applicable
Sample
Log
Results
System Status
Probable Cause
Host/ToDo
N/A
N/A
Host sent new order for teh specified IDs
Host/ToDo
N/A
N/A
Host sent new demographic for the existing patient ID.
Calibration
N/A
N/A
When no lot number has been entered for calibration
table.
Calibration
Discarded
Ready
Operator ran a latex control sample and no latex control
file was set up.
Event
Sample run
The Workstation
The sample run received does not match a ToDo List
status set to
receives the number
entry. This could indicate:
NO MATCH.
of consecutive NO
Results
MATCH errors shown
reported.
in the After # Runs
field. An Operator
Alert window also
appears, and the
Analytical Station
stops processing
samples.
Event
Sample run
The Workstation
The sample run received is not properly identified. This
status set to
receives the number
condition can occur only when the positive
NO READ.
of consecutive NO
identification scheme used is Cass/Pos and SID.This
Results
READ errors shown
could indicate:
reported.
in the After # Runs
field. An Operator
Alert window also
appears, and the
Analytical Station
stops processing
samples.
System Status/Recovery
Corrective Action
Status only: no action necessary.
See Host Transmission Specification for correct format.
Enter lot number and save.
Create a latex control file and rerun the control.
1. Select
r Preassigned Cass/Pos or sample ID was entered
2. Check the ToDo list. If identification information is incorrect, edit the
incorrectly.
r Sample was run in incorrect test mode
3. Check the cassette, tube, and labels. Ensure that the tube is in the correct
r Tube was misplaced in the cassette
r Incorrect test assigned in the ToDo List.
4. Clean the bar-code reader.
r Label is smudged.
5. If the message persists, contact your Beckman Coulter Representative.
r Label is misplaced.
r Bar-code reader is dirty.
1. Select
2. Check the cassette, tube, and labels. Ensure that the tube is in the correct
r Label is smudged.
r Tube or label is misplaced.
3. Clean the bar-code reader.
r Bar-code reader is dirty.
4. If the message persists, contact your Beckman Coulter Representative.
6-62
on the Operator Alert window.
information.
position and the label is properly placed and readable. If necessary,
reposition the tube or label, or replace the label for future use.
on the Operator Alert window.
position and the label is properly placed and readable. If necessary,
reposition the tube or label, or replace the label for future use.
PN 773023AC