Beckman Coulter LH 780 Instructions Manual page 217

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Table 6.18 Workstation Messages - C through G (Continued)
Workstation
Log Message and Icon on Command Center
Control run for control folder <Lot #> out of range.
Yellow traffic light is displayed if Control AutoStop option is
disabled.
Red traffic light is displayed if Control AutoStop option is
enabled.
Control run outside limits condition has cleared
Daily check precision test failed: (<List Of Parameters That
Failed With Result>).
Daily check ramp test failed: (<List Of Parameters That Failed
With Result>).
Demographic information for<Patient ID> updated by Host
with order [<Sample ID>,<CassPos>](<Last Name>, <First
Name>)
Duplicate run date and/or run time received (<Run SID>,
<Run C/P>).
Flow cell voltages/pressures out of specifications (<Run
SID>, <Run C/P>).
PN 773023AC
Applicable
Sample
Log
Results
System Status
Probable Cause
Control
N/A
Analytical Station
At least one of the parameter results for the control run
stopped if Control
received is outside the expected range (or lab limits).
AutoStop option is
This could indicate:
enabled.
Note: The AutoStop
option only applies to
controls analyzed in
the Automatic Mode.
Controls
N/A
Ready
The previous control run was outside the limits, the
current control run is within the limits.
Daily Checks
N/A
Ready
During a Daily Check operation (StartUp), one of the
Precision test parameters failed
Daily Checks
N/A
Ready
During a Daily Check operation (StartUp), one of the
Ramp test parameters failed.
Host / ToDo
N/A
N/A
Host sent new demographic information for the existing
List
Patient ID.
Event
N/A
Analytical Station
The LH Workstation received a sample run with an
stopped.
invalid run date and/or run time or two sample runs
with the same date and time.
Instrument
"System
Ready
Analytical Station detected a flow cell problem such as
Alarm"
Diff or Retic sheath pressure or voltages out of
message
specifications.
printed in
footer of
sample
report.
System Status/Recovery
Corrective Action
Refer to the procedure in the Online Help System and in Chapter 3 of the
Operator's Guide for what to do when a control is outside its expected ranges or
what to do when a latex control is outside its expected ranges.
r Improper mixing.
r Control is stored in the wrong folder.
r Control information was set up incorrectly.
r Statistical outlier.
r Change in the control material.
r Bubbles or clog in the flow cell (latex controls only).
r Analytical Station problem.
Status only: no action necessary.
1. Verify that the baths contain only clean diluent (no bleach or cleaning agent)
2. Repeat the Precision test at the Analyzer.
3. If the message recurs, call your Beckman Coulter Representative.
1. Verify that the baths contain only clean diluent (no bleach or cleaning agent)
2. At the Analyzer, repeat the Ramp test.
3. If the message recurs, call your Beckman Coulter Representative.
Status only: no action necessary
1. Check the clock on the Analyzer screen to see if it is incrementing.
2. If the clock is not running, reset the Analytical Station.
3. If the clock is still not running, call your Beckman Coulter Representative.
If message persists, call your Beckman Coulter Representative.
TROUBLESHOOTING
WORKSTATION MESSAGES
and that the diluent covers the apertures.
and that the diluent covers the apertures.