Solution, - Beckman Coulter LH 780 Instructions Manual

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Dispensing Retic Clearing Solution
1.
Go to the Numeric Keypad.
2.
Press
chamber. The Numeric Keypad displays RETIC CLEAR while the instrument performs
this function. When the function completes processing, the Numeric Keypad displays
FUNCTION = 26. Press
3.
Press
Performing Sheath Tank Check
1.
Go to the Numeric Keypad.
2.
Press
Drain sheath tank
r
Fill sheath tank
r
Check sheath tank level
r
3.
The Numeric Keypad displays SHEATH TANK CHECK while the instrument performs
this function. When the function completes processing, the Numeric Keypad displays
FUNCTION = 20. Press
4.
Press
Freeing Solenoids
1.
Go to the Numeric Keypad.
2.
Press
Keypad displays SOLENOIDS FREE while the instrument performs this function. When
PN 773023AC
E
N
T
E
2
6
F
R
to dispense Retic clearing solution into the mixing
E
N
T
E
R
to perform this function again.
STOP
to exit this function. The Numeric Keypad displays READY.