Error Messages, - Beckman Coulter LH 780 Instructions Manual

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

6.1
SYSTEM AND WORKSTATION MESSAGES
The LH 700 SERIES System automatically monitors many of its activities and conditions and
generates messages to alert the operator when the system does not respond as expected.
Messages that apply to the LH 700 SERIES System are described in the tables in
ANALYTICAL STATION
described in the tables in
referenced in the index. Any messages that are logged at the Workstation appear in the
following format: date, time, instrument name, operator name, message.
When an event causes the system to stop processing samples, double-click on
the Status dialog box. Place the cursor over an entry in the Status box to display additional
information. Double-click on an entry in the Status box if you want to display the error
message help topic for that event. Select
processing.
Note: Not all system events require corrective action. Error message help topics are displayed
only for the system events that have corrective actions that you can perform.
System Alarms
Many of the System error messages are accompanied by an alarm. If the alarm is a
r
Diluter-generated alarm, press
If the alarm is generated in the SlideMaker, press any key on the SlideMaker screen.
r
About Table 6.2, System messages
The System messages, with their probable causes and corrective actions, are arranged
alphabetically by Analyzer message.
Several System messages are listed with an "and" condition. For those messages:
If the error occurs while the instrument is idle or ready to pierce, the Analyzer only
r
displays the first message.
If the error occurs while the instrument is running, the Analyzer displays the first
r
message and then replaces it with the second message. The second message is always
Diluter Error or Non-Diluter Error.
To locate the appropriate corrective actions in the System Message tables for Diluter Error
and Non-Diluter Error messages, use the message on the Diluter screen; it usually
mirrors the original Analyzer message.
When a message column lists "current status" instead of a specific message, one of the
following Analyzer status messages is displayed:
ACTIVE
ANALYZE
PN 773023AC
MESSAGES; messages that apply to the Workstation only are
Heading 6.5, WORKSTATION
to acknowledge the event and continue
.
ALARM RESET
:
LH 700 SERIES
INITIALIZING/RAW
INTRO SAMPLE
6
ERROR MESSAGES
Heading 6.4,
MESSAGES. All the messages are
READ HGB
READY
6
to display
6-1