TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Table 6.13 System Messages - S (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Stack Overflow
NOT READY
N/A
Stain Heater Disabled
(current status)
STAIN HEATER
DISABLED
Stain Heater Sensor Bad (current status)
STAIN HEATER
SENSOR BAD
Stain Temp Hi. Heater
(current status)
STAIN TEMP. HI
Disabled
HEATER DISABLED
Stain Temp Low. Run
(current status)
STAIN TEMP LOW.
Stopped
RUN STOPPED
Stripper Plate Not In
SYSTEM FAULT
STRIPPER PLATE
NOT IN
and
MANUAL MODE
6-44
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
No
Instrument Log
Data displayed if available
(depends upon when error
Internal software failure.
occurred).
Reset Analytical Station.
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Stain heater disabled for
Alarm" message printed in
current cycle. Heater
footer of sample report
temperature not within
(depends upon when error
expected limits. Reset
occurred).
Analytical Station.
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Stain temperature sensor
Alarm" message printed in
failed. Reset Analytical
footer of sample report
Station.
(depends upon when error
occurred).
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Stain temperature high,
Alarm" message printed in
heater disabled. Reset
footer of sample report
Analytical Station.
(depends upon when error
occurred).
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Stain temperature low,
Alarm" message printed in
heater disabled. Reset
footer of sample report
Analytical Station.
(depends upon when error
occurred).
Yes
Instrument Log
Data displayed if available
(depends upon when error
Stripper plate obstructed
occurred).
(not in) or faulty sensor.
Remove obstruction and
restart Analytical Station.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Stopped - inoperable.
Internal software problem.
Run stopped when Retic
Temperature control failure.
cycle is run.
If this error occurs it will
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
Run stopped when Retic
Faulty or disconnected sensor.
cycle is run.
If this error occurs it will
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
Run stopped when Retic
Temperature control failure.
cycle is run.
If this error occurs it will
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
Run stopped when Retic
Temperature control failure.
cycle is run.
If this error occurs it will
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
Automatic mode disabled;
Obstruction or faulty sensor.
Manual mode operable.
6-44
Corrective Action
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Ensure the stripper plate is unobstructed.
2. Reset the Analytical Station.
3. Continue normal operation. If the message recurs,
call your Beckman Coulter Representative. You can
run samples in the Manual mode.
PN 773023AC