CLEANING PROCEDURES
BLEACHING RBC INTERNAL ELECTRODES
B Put the Bleach Solution in the WBC Bath
C Put the Bleach Solution in the RBC Bath
2-28
1/2 in.
1/2 in.
1.
Open the Diluter door.
2.
Ensure pneumatics are on.
3.
Go to the Numeric Keypad.
DRAIN
4.
Press
until the aperture baths
drain completely into the waste
chamber.
5.
Find the bleach fitting check valve on
the top right of the WBC bath.
6.
Connect the loose end of the tubing
on the wash bottle to the bleach
check valve on the WBC bath.
7.
Put bleach solution into WBC bath so
solution is within 1/2 in. from the top
of the bath
1.
Disconnect the bleach tubing from
the check valve on the top of WBC
bath and connect it to the bleach
check valve on the top right of the
RBC bath.
2.
Put bleach solution into the RBC bath
so the solution is within 1/2 in. from
the top of the bath.
3.
Disconnect the bleach tubing from
the RBC bath and reconnect it to the
WBC bath. You may need to refill the
WBC bath. Monitor the bleach level
in the bath to ensure the liquid level
remains above the apertures.
PN 773023AC