D Use Distilled Water to Clean the Bath
PN 773023AC
CLEANING PROCEDURES
CLEANING AN APERTURE BATH
1.
Moisten a lint-free tissue with
distilled water.
2.
Use the moist lint-free tissue to
clean the openings in the bath so
they are free of salt deposits.
IMPORTANT
Risk of instrument damage. If
you touch the center of the block you can
damage the instrument. Do not touch the
aperture in the center of the block.
3.
Look at each large O-ring to ensure
it is free of salt deposits.
4.
If necessary, use the moist lint-free
tissue to clean the area on and
around the three large O-rings.
5.
Replace the O-rings if necessary.
2
2-11