CLEANING PROCEDURES
CLEANING AN APERTURE BATH
Procedure
DRAIN
A Press
POWER OFF
B Press
C Unhook Springs and Pull Out Bath
2-10
until the Aperture Baths Drain Completely into the Waste Chamber
1.
Open the Diluter.
2.
Manually move the rocker bed to
its backward position.
3.
Use a hemostat to unhook the
spring attached to each side of the
bath.
IMPORTANT
Risk of incorrect control
results. If you move the external electrode
attached to the bath you can create a
carryover problem. Do not move the
external electrode.
4.
Using both hands, gently rock the
bath off the large O-rings.