TROUBLESHOOTING
COMMUNICATION HAS BEEN LOST BETWEEN THE ANALYTICAL STATION AND WORKSTATION
AutoStop Options
Category
Control
NO MATCH
NO READ
Partial Aspiration
SlideMaker
Voteout
XB
XM
5.16 COMMUNICATION HAS BEEN LOST BETWEEN THE ANALYTICAL STATION AND
WORKSTATION
There are situations when the Analytical Station and the Workstation will stop
communicating with one another. The most effective way to restore communication is to
Shutdown and restart the Workstation. If this is not successful to restore communication, you
may be required to shutdown the Workstation, reset the Analytical Workstation, restart and
then log on to the Workstation. It is not imperative to log on the Workstation but it must be
restarted prior to the completion of the Analytical Station reset in order to avoid another
message.
Note: Periodic optimization procedures will help the function of the Workstation. By shutting
down the Workstation daily while performing the Analytical Station Shutdown, you can
5-42
Definition
A control failure has occurred. Detail is
provided in Control AutoStop On.
Applies to both traditional Controls and
Extended QC.
The Workstation can not find a match
of the positive sample identifier to the
preassigned ID in the To Do list.
The Analytical Station cannot read a
bar-code label and sample ID is used as
a positive ID.
The Analytical Station encountered,
partial aspiration error disabled blood
detectors blocked tubing or needle, or
insufficient sample volume.
A SlideMaker failure has occurred.
Sample results contain a parameter
with a Voteout.
Batch values exceed the defined
percent differences from XB Target
values.
Batch values exceed the defined
percent differences from XM Limits.
Configuration Options
Disabled or Enabled for all.
Disabled or
Enabled: 1 – 10. When consecutive
runs meet the configured threshold, the
system will halt.
Disabled or
Enabled: 1 – 10. When consecutive
runs meet the configured threshold, the
system will halt.
Disabled or
Enabled: 1 – 10. When consecutive
runs meet the configured threshold, the
system will halt.
Disabled or Enabled for all.
Disabled or
Enabled: 1 – 10. When consecutive
runs meet the configured threshold, the
system will halt.
Disabled or
Enabled: 1 or 2. When consecutive
batches meet the configured threshold,
the system will halt.
Disabled or
Enabled: 1 or 2. When consecutive
batches meet the configured threshold,
the system will halt.
PN 773023AC