Table 6.9 System Messages - L (Continued)
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Level Check Fail Caused
(current status)
LEVEL CHECK FAIL
Stop
CAUSED STOP
Level -D-----
READY
LEVEL -D-----
Level --L----
READY
LEVEL --L----
Level ---P---
READY
LEVEL ---P---
Level ----R--
READY
LEVEL ----R--
Level -----V-
READY
LEVEL -----V-
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data displayed if available
(depends upon when error
Replace reagent container.
occurred).
Yes
Instrument Log
Data displayed if available
(depends upon when error
Diluent is low. Replace
occurred).
diluent container and
press START/CONT on
Numeric Keypad to
continue.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Reservoir low on CBC
occurred).
lyse. Replace CBC Lyse
container and press
START/CONT on Numeric
Keypad to continue.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Reservoir low or Stabilyse
occurred).
level sense detected.
Replace DIFF PAK or
Stabilyse container and
press START/CONT on
Numeric Keypad to
continue.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Reservoir low or Stain or
occurred).
clearing agent level sense
detected. Replace Retic
PAK or Stain or clearing
agent container and press
START/CONT on Numeric
Keypad to continue.
Yes
Instrument Log
Data displayed if available
(depends upon when error
Vacuum overflow tank
occurred).
full. Empty vacuum
overflow tank and press
START/CONT on Numeric
Keypad to continue.
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Run stopped.
Analytical Station detected a
low fluidic condition that could
If this error occurs it will
produce a level sense problem.
not halt the action in
process, but it will prevent
initiation of a new sample,
unless the error condition
is corrected.
Cycling inhibited until
Diluent is low.
sensor error resolved.
Cycling inhibited until
Reservoir low on CBC lytic
sensor error resolved.
reagent.
Cycling inhibited until
Reservoir low on diff lytic
sensor error resolved.
reagent and/or diff preservative.
Cycling inhibited until
Reservoir low on retic stain
sensor error resolved.
and/or retic clearing solution.
Cycling inhibited until
Vacuum overflow tank full.
sensor error resolved.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Check the reagent containers and replace any that
are low.
2. Press START/CONT to resume normal operation.
3. If the message recurs, call your Beckman Coulter
Representative.
1. Replace diluent container.
2. Press START/CONT to resume normal operation.
1. Replace CBC lytic reagent container.
2. Press START/CONT to resume normal operation.
1. Replace DIFF PAK or Stabilyse container.
2. Press START/CONT to resume normal operation.
1. Replace RETIC PAK or Stain container.
2. Press START/CONT to resume normal operation.
1. Empty vacuum overflow tank.
2. Press START/CONT to resume normal operation.
6-31