Beckman Coulter LH 780 Instructions Manual page 225

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Table 6.21 Workstation Messages - L through M (Continued)
Workstation
Log Message and Icon on Command Center
Last two Diff XM batches are outside limits.
Yellow traffic light is displayed if XM AutoStop option is
disabled.
Red traffic light is displayed if XM AutoStop option is
enabled.
Last two Retic Calculated XM batches are outside limits.
Yellow traffic light is displayed if XM AutoStop option is
disabled.
Red traffic light is displayed if XM AutoStop option is
enabled.
Last two Retic XM batches are outside limits.
Yellow traffic light is displayed if XM AutoStop option is
disabled.
Red traffic light is displayed if XM AutoStop option is
enabled.
Manual CBC calibration factors updated from LH Workstation
Maximum # Of Repro runs Exceeded - Sample Not Stored
PN 773023AC
Applicable
Sample
Log
Results
System Status
Probable Cause
XM
N/A
Analytical Station
The last XM batch is outside limits.
stopped if XM
AutoStop option (2
batches out) is
enabled.
XM
N/A
Analytical Station
The last XM batch is outside limits.
stopped if XM
AutoStop option (2
batches out) is
enabled.
XM
N/A
Analytical Station
The last XM batch is outside limits.
stopped if XM
AutoStop option (2
batches out) is
enabled.
Calibration
N/A
Ready
Secondary calibration factor was changed.
Reproducibilit
Discarded
Ready
User attempted to run more than the maximum
y
allowable reproducibility cycles.
System Status/Recovery
Corrective Action
1. Check the results for a nonrandom sample distribution.
r If a nonrandom sample distribution is the problem, document the problem based
on your laboratory's protocol and continue processing samples.
r Otherwise, troubleshoot for a possible Analytical Station problem.
1. Check the results for a nonrandom sample distribution.
r If a nonrandom sample distribution is the problem, document the problem based
on your laboratory's protocol and continue processing samples.
r Otherwise, troubleshoot for a possible Analytical Station problem.
1. Check the results for a nonrandom sample distribution.
r If a nonrandom sample distribution is the problem, document the problem based
on your laboratory's protocol and continue processing samples.
r Otherwise, troubleshoot for a possible Analytical Station problem.
Status only: no action necessary.
Change operating mode from Reproducibility to Auto Analysis
6
TROUBLESHOOTING
WORKSTATION MESSAGES
6-61