Cleaning Apertures, - Beckman Coulter LH 780 Instructions Manual

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

F27 - Dispensing CBC Lytic Reagent
1.
Go to the Numeric Keypad.
2.
Press
Note: This function is intended for troubleshooting, not for priming CBC lytic reagent.
The Numeric Keypad displays CBC LYSE while the instrument performs this function.
When the function completes processing, the Numeric Keypad displays FUNCTION =
27. Press
3.
Press
displays READY.
Cleaning Apertures
1.
Go to the Numeric Keypad.
2.
Press
protein or debris from the apertures. The Numeric Keypad displays CLEAN APERTURES
while the instrument performs this function. When the function completes processing,
the Numeric Keypad displays FUNCTION = 01. Press
again.
3.
Press
Running Latex Control—Diff and Retic
1.
Perform instrument startup.
2.
Check that the instrument process type on the Command Center is set to
3.
Ensure the latex primer and control are within the correct temperature range. For
COULTER LATRON primer and control the correct temperature range is
18-30°C/64-86°F .
PN 773023AC
E
N
T
E
2
F
7
R
to dispense 1.04 mL CBC lytic reagent into the WBC bath.
E
N
T
E
R
to perform this function again.
STOP
to exit this function. The RBC and WBC baths drain. The Numeric Keypad
E
N
T
E
0
F
1
R
to apply current across the apertures. The current removes
STOP
to exit this function. The Numeric Keypad displays READY.
TROUBLESHOOTING
DILUTER FUNCTIONS
E
N
T
E
R
to perform this function