TROUBLESHOOTING
RUNNING BACKGROUND TEST
Table 5.1 Solenoid Functions (Continued)
111
Probe Wipe Vacuum
112
Not used
113
Not used
114
Needle Vent Side Isolator (SlideMaker
Option)
115
VC19 Isolator 1 (SlideMaker Option)
116
VC19 Isolator 2
117
SlideMaker Aspiration Line Isolator (SM)
118
Reservoir SW1 (SM) (N.O. RES1)
119
Not used
120
Duplicate SlideMaker Aspiration Line
Isolator (SM) (Duplicate of SL117)
121
VLS Check
122
Sweep Flow Reservoir Refill
123
Probe Wipe Block Up
124
Probe Wipe Block Reset
125
Probe Wipe Rinse Cup
126
Segment Diff Blood Sample
127
Segment Retic Blood Sample
128
Blood/Stain Sample Segment Valve Return
129
Ghost Line Protection (inverted signal)
131
5 PSI, WBC Bath
5.4
RUNNING BACKGROUND TEST
The Background test includes Hgb voltage check, Diff test, and Retic test.
1.

Go to the Analyzer screen.

2.
Press
MAIN MENU
3.
Press
ANALYZER FUNCTIONS
4.
Press
START UP TESTS
5.
Press
BACKGROUND TEST
pulses. After showing the pulses, the Analyzer displays the
transmits the test results to the Workstation. The Analyzer screen then returns to the
SYSTEM RUN
6.
Review the results at the Workstation. If results are unacceptable, rerun the background
test. If the results are unacceptable on the second try, call your Beckman Coulter
Representative.
7.
If background test results are acceptable, begin running controls or press
display the Analyzer
5-32
.
.
.
. The Analyzer screen displays a
screen.
MAIN MENU
screen.
VL111
VL114
VL115
VL116
VL117
VL118
VL120
Logic Signal to Trigger VLS Check
Logic Signal to Trigger Sweep Flow Reservoir Refill
Logic Signal to Trigger Probe Wipe Block Up
Logic Signal to Trigger Probe Wipe Block Reset
CL11
CVL85/126 (Future component/function)
CVL87/127 (Future component/function)
CVL93/128 (Future component/function)
VL129
VL131864
screen that shows
SYSTEM RUN
START UP TESTS
screen and
MAIN MENU
to
PN 773023AC

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents