Beckman Coulter LH 780 Instructions Manual page 211

Hematology analyzer
Hide thumbs Also See for LH 780:
Table of Contents

Advertisement

Table 6.15 System Messages - U
Analytical Station
Analyzer
Diluter
Analyzer Alert Message
Status Message
Screen Message
Unknown Diff Error
(current status)
N/A
Message
UNKNOWN Diluter Error
(current status)
N/A
Message
Note: If event severe,
this message may not
appear.
Vls Air Error
MANUAL MODE
VLS AIR ERROR
Vls Diluent Error
MANUAL MODE
VLS DILUENT
ERROr
PN 773023AC
Workstation
Beep
Log Message and
s
Icon on Command Center
Sample Results
Yes
Instrument Log
Data displayed.
AMC-DIFF communication
failure: unknown. Reset
Analytical Station.
Yes
Instrument Log
Data displayed.
Analyzer-Diluter
communication failure:
unknown. Reset Analytical
Station.
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Vls Air Error
Alarm" message printed in
footer of sample report
(depends upon when error
occurred).
Yes
Instrument Log
Data not transmitted, or data
transmitted and "System
Vls Diluent Error
Alarm" message printed in
footer of sample report
(depends upon when error
occurred).
Analytical Station Status/Recovery
Analytical Station Status
Probable Cause
Idle - ready for request.
Communication error between
DIFF PROCESSOR card and
AMC.
Idle - ready for request.
AMC detected invalid error code
transmitted by Diluter.
Automatic mode is
This message displays when a
disabled. Manual mode is
problem is encountered with
operable.
the vent line.
Automatic mode is
Pinched, kinked or cut tubing.
disabled. Manual mode is
operable.
TROUBLESHOOTING
ANALYTICAL STATION MESSAGES
Corrective Action
1. Reset the Analytical Station.
2. Continue normal operation and watch for
messages.
3. If additional messages appear, follow the corrective
actions associated with the messages.
4. If this messages persists:
a. Select the CBC only test mode.
b. Run CBC only samples.
c. Call your Beckman Coulter Representative.
1. Reset the Analytical Station.
2. Continue normal operation and watch for
messages.
3. If additional messages appear, follow the corrective
actions associated with the messages.
4. If this message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
1. Reset the Analytical Station.
2. If message persists, call your Beckman Coulter
Representative.
6-47
6