Resetting The Compliance Film Calibration, Internal Densitometer Calibration And Thermal Print Head Calibration; Performing Internal Densitometer Calibration, Film Calibrations And Thermal Print Head Calibration - AGFA DRYSTAR 5302 Technical Documentation Manual

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
2.5.3
Resetting the Compliance Film Calibration, Internal Densitometer Calibration and
Thermal Print Head Calibration
NOTE:
Resetting the Compliance Film Calibration, Internal Densitometer Calibration and
Thermal Print Head Calibration is required to be sure, that possibly wrong calibrations
from previous service interventions are deleted.
(1) In the service command session select "6 Calibration – 6 Compliance film
calibration – 1 Reset compliance film calibration for upper tray.
(2) Select "2 Reset compliance film calibration for lower tray".
(3) Leave the "service command session".
(4) In the service tools page select "Reset calibration data".
(5) Select "Default CDM" and "Default Profile". Note, that "Default Profile" button is not
available, if no "Print Head Profile" has been performed in the field yet.
(6) Leave the Reset calibration data" screen.
NOTE:
After this action the printer will show following warnings:
HC302016 Lower tray needs film calibration
HC301016 Upper tray needs film calibration
HC306014 Printer needs CDM calibration
This problem will be solved in the next section.
2.5.4
Performing Internal Densitometer Calibration, Film Calibrations and Thermal
Print Head Calibration
NOTE:
The Internal Densitometer Calibration, Film Calibration and Thermal Print Head
Calibration is described in this chapter (3.6 Adjustments and Calibrations) in detail.
Here only a short description is given. For details refer to:
2.2 CDM (Internal Densitometer) Calibration
2.1 Film Calibration
2.3 Print Head (TH) Calibration
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Chapter 3.6 / 21
Agfa Company Confidential