64 | Drystar AXYS XL | Advanced Operation (key-operator mode)

Film calibration

The film calibration combines Dmax and sensitometry calibration, and is
activated as one single procedure. Separate Dmax and sensitometry
calibration steps are not required.
1. On the Select calibration menu, select 'Film calibration'.
The 'Select tray' screen appears:
2. Press the Up/Down arrow keys to select the desired tray, followed by the
Confirm key.
The 'Film calibration' screen appears:
The counter indicates the progress of the calibration process. The
calibration film is measured automatically by the built-in MDM (Macro
Densitometer). The printer is being calibrated during the printing process.
Press the Confirm key to retry calibrating.
3. If the film calibration failed, the following message will appear.
Press the Confirm key to retry return to the calibration menu.
4. When the film calibration has been completed successfully, three cases are
possible:
• Calibration successful, results OK.
3753B EN 20210520 0952
Note: In this part, we always will elaborate a case for the lower
tray. The procedure of film calibration is identical for the upper
tray.
Note: If the printer cannot read the Film identification tag,
no calibration is possible and the following message is
displayed: