Additional Explanation (Data Logging)

When 9 or More Data Items Are Output as One Record
Up to 8 Value can be output with one [Data Logging] Processing Item. When 9 or more data items are to be
output as 1 record, perform settings in the following manner.
• Registers two or more [Data Logging] units in one scene. (*1)
• Set [File name] of [Data Logging] so that it is identical. (*2)
• This prepares set up to attach "Record separator (CR+LF)" to the end of all data output. (*3)
Example) When outputting the coordinate data for 12 points acquired in two "Search" of
measurements performed on substrate arrangement in 1 record.
0. Camera Image Input
1. Search
(Search position X, Y, Search angle TH, Reference SX, SY, Ref. Angle ST)
2. Search
(Search position X, Y, Search angle TH, Reference SX, SY, Ref. Angle ST)
3. Data logging
4. Data logging
Output record:
YYYY-MM-DD_HH-MM-SS-MS,X,Y,TH,SX,SY,ST,X,Y,TH,SX,SY,ST+CR+LF
Unit 3 [Data logging]
setting details
<Condition setting>
Output Destination
datalog.csv
(File name)
Integer
8
Decimal
3
Measurement ID
ON
Minus
-
0 suppress
OFF
Field separator
Comma
Record separator
Comma
<Output data>
Calculation 0. U1.X (Search position X)
Calculation 1. U1.Y (Search position Y)
Calculation 2. U1.TH (Angle θ)
Calculation 3. U1.SX (reference X)
Calculation 4. U1.SX (reference Y)
Calculation 5. U1.ST (Reference angle θ)
Calculation 6. U2.X (Search position X)
Calculation 7. U2.Y (Search position Y)
FH/FZ5 Processing Item Function Reference Manual
(*1)
1 Set up more than "data logging" Unit in one scene
Data of Unit 1
Data of Unit 2
Unit 4 [Data logging]
setting details
<Condition setting>
Output Destination
(File name)
Integer
Decimal
Measurement ID
Minus
0 suppress
Field separator
Record separator
<Output data>
Calculation 0. U2.TH (Search angle θ)
Calculation 1. U2.SX (reference X)
Calculation 2. U2.SY (reference Y)
Calculation 3. U2.ST (Reference angle θ)
(*2)
datalog.csv
Make the path and file name the
same.
8
3
OFF
-
OFF
Comma
(*3)
CR+LF
Set "Record separator (CR+LF)" in
unit 4 which contains the last data
The data not included in Unit 3 will be
output as Calculation 0 to 3 in Unit 4
Remarks
687
Data Logging
4