Frequency Limits - Agilent Technologies E1445A User's Manual And Scpi Programming Manual

Arbitrary function generator module scpi programming guide
Hide thumbs Also See for E1445A:
Table of Contents

Advertisement

Frequency Limits

Function
DC
SINusoid
SQUare
TRIangle
RAMP
USER
* Multiply High Limit frequency values by 2 if frequency doubling is selected by the
[SOURce:]FREQuency[1]:RANGe command.
470
Useful Tables
Table B-3. Agilent E1445A Frequency Limits
Trigger Start
Source
N/A
INTernal[1]
INTernal[1]*
INTernal2
INTernal[1]*
INTernal2
INTernal[1]*
INTernal2
INTernal[1]*
INTernal2
Frequency
Low Limit
High Limit
N/A
0 Hz
Ref Osc freq/4
0 Hz
Ref Osc freq/16
Ref Osc freq/4/
Ref Osc freq/4
131072
0 Hz
Ref Osc freq/
Ramp Points
Ref Osc freq/
Ref Osc freq/
131072/
Ramp Points
Ramp Points
0 Hz
Ref Osc freq/
Ramp Points
Ref Osc freq/
Ref Osc freq/
131072/
Ramp Points
Ramp Points
0 Hz
Ref Osc freq/4
Ref Osc freq/
Ref Osc freq
131072
N/A
4/
4/
Appendix B

Advertisement

Table of Contents
loading