:SECure[:STATe]

Parameters
Comments
Example
Chapter 8
CALibration:SECure[:STATe] <
security. Calibration security must be disabled to calibrate the Agilent E1445A,
read or write calibration data, change the security code, or change the protected user
data.
Parameter
Parameter
Name
<
mode
>
boolean
<
code
>
Attempting to disable calibration security without providing the
generates Error -109,"Missing parameter". The value supplied must match the
currently programmed security code or Error -224,"Illegal parameter value" will
be generated. The Agilent E1445A will then wait 1 second before executing any
subsequent commands.
To enable security, the
present.
Executable when Initiated: Yes
Coupling Group: None
Related Commands:
CALibration[:DC]:BEGin
CALibration:DATA:AC2
*RST Condition: Unaffected
Power-On Condition:
Disabling Calibration Security
CAL:SEC:STATe OFF,E1445A
mode
>[,<
code
>]
Range of
Type
Values
OFF | 0 | ON | 1
string
0 through 15 characters
code
parameter is not not required, but is checked if it is
<
>
,
,
CALibration:SECure:CODE
CALibration:SECure[:STATe] ON
Disables security assuming factory-set
security code.
CALibration
enables or disables calibration
Default
Units
none
none
<
code
>
parameter
CALibration:DATA:AC[1]
,
*PUD
Command Reference
,
303