Activation
One electrode
One linear control ele­
ment
Program 5: VarioDual
Coding plug, yellow: 13E184=5
Activation
Two electrodes
Program 6: DMC plus Sensorik
Coding plug, purple: 13E184=6
Activation
Two electrodes
6.3.2 Program description
INFORMATION
Both the automatic gripping force readjustment and the FlexiGrip function can be stopped at
any time with a brief "OPEN" impulse.
INFORMATION
All gripping force specifications (e.g. 100 N ~ 10 Kp) are non-binding guiding values to illustrate
the functional principle of the various operating modes.
6.3.2.1 Program 1: DMC plus Sensorik
Control with two electrodes
This control variant corresponds to DMC plus control with the integrated "virtual hand switch"
(dynamic mode control) but also features the "SUVA sensors" grip stabilisation system. The level
of the gripping speed or gripping force is determined by the strength of the electrode signal (res­
ulting from the muscle contraction). After gripping with maximum force, the activation threshold in
the OPEN direction is raised to a higher value ("virtual hand switch"). This increase reduces the
34
Open
Speed and level of
muscle contraction on
the electrode
Speed:
Proportional
Speed and level of pull
on the linear control
element
Speed:
Proportional
Open
Speed and level of
muscle contraction on
the first electrode
Speed:
Proportional
Open
Muscle signal via the
electrode
Speed:
Proportional
Close
Speed and level of
muscle relaxation on
the electrode
Speed:
Proportional
Speed of relaxing the
pull on the linear con­
trol element
Speed:
Proportional
Close
Speed and level of
muscle relaxation on
the first electrode
Speed:
Proportional
Gripping force propor­
tional to strength of
the muscle signal on
the second electrode
Close
Muscle signal via the
electrode
Speed:
Proportional
Indication
For patients with one
muscle and a strong
muscle signal or tend­
ency to co-contraction
For patients with
muscle signal that is
too weak or no muscle
signal
Indication
For patients with two
strong muscle signals
Indication
For patients with two
strong muscle signals
SUVA sensors and
FlexiGrip can be deac­
tivated