6.3.7
High MV
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Setting range
Setting resolution
Ventilator action
118
Alarms
User manual
Alarm (High Inspired Minute Volume Alarm)
i
Description
High MVi
Medium
A High Inspired Minute Volume alarm will be given when
the monitored inspired minute volume exceeds the set limit
for 15 seconds.
Mismatch between Breath Rate, Inspired Tidal Volume set-
tings and the alarm setting.
Increased Breath Rate.
Leakage around the mask or within one of the components
of the circuit.
When inspired minute volume is below set alarm limits
Adult mode: 1.0 to 40 l/min
Pediatric mode: 1.0 to 20 l/min
Off
0.5 l/min
The ventilator will continue treatment with the same settings.
Doc. 007301 K-1