6.3.7
High MV
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Setting range
Setting resolution
Ventilator action
118
Alarms
User manual
Alarm (High Inspired Minute Volume Alarm)
i
Description
High MVi
Medium
A High Inspired Minute Volume alarm will be given when
the monitored inspired minute volume exceeds the set limit
for 15 seconds.
•
Mismatch between Breath Rate, Inspired Tidal Volume set-
tings and the alarm setting.
•
Increased Breath Rate.
•
Leakage around the mask or within one of the components
of the circuit.
When inspired minute volume is below set alarm limits
•
Adult mode: 1.0 to 40 l/min
•
Pediatric mode: 1.0 to 20 l/min
•
Off
0.5 l/min
The ventilator will continue treatment with the same settings.
Doc. 007301 K-1