6.4.10
Poor SpO
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
Reset
6.4.11
CO
Disconnected (CO
2
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
Reset
142
Alarms
User manual
Signal
2
Description
Poor SPO2 Signal
A Poor SpO
nal is not correct. Check the SpO
High
Artifact or low perfusion reported by SpO
The ventilator will continue treatment with the same settings.
OK message from SpO
connected by user or SpO
Sensor Failure/Disconnection Alarm)
2
Description
CO2 Sensor Disconnected
A CO
Sensor Failure/Disconnection alarm will be given
2
when communication between the ventilator and the CO
sensor has been lost for 2 seconds. Check the CO
High
•
CO
Sensor disconnected.
2
•
Failure in the CO
The ventilator will continue treatment with the same settings.
Confirmation by user or reconnected/changed.
signal alarm will be given when the SpO
2
sensor.
2
electronics or SpO
2
Signal Lost alarm is triggered.
2
sensor.
2
sig-
2
electronics
2
electronics dis-
2
2
sensor.
2
Doc. 007301 K-1