5.5.4
Breath Rate
Item
Definition
Setting max
Setting
resolution
Default
value
5.5.5
Backup Rate
Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
80
How to Use the Vivo 45 LS
User manual
Description
The Breath Rate setting defines the minimum number of breaths the
ventilator will deliver as long as no inspiratory trigger effort from the
patient is detected. The cycles will be ventilator-initiated breaths.
The combination of the Breath Rate and Inspiratory Time setting is
limited by the I:E ratio range 1:9.9 to 2:1.
40 bpm (Adult), 60 bpm (Pediatric mode)
1 bpm
12 bpm (Adult mode)
20 bpm (Pediatric mode)
Description
The Backup Rate setting defines the minimum number of breaths the
ventilator will deliver as long as no inspiratory trigger effort from the
patient is detected. The cycles will be ventilator-initiated breaths.
The combination of the Backup Rate and Inspiratory Time setting is
limited by the I:E ratio range 1:9.9 to 2:1 (1:99 to 1:1 in MPV).
4 bpm (Adult)
6 bpm (Pediatric mode)
0 bpm (MPV modes)
30 bpm (Adult), 30 bpm (MPV), 60 bpm (Pediatric mode)
1 bpm
12 bpm (Adult mode)
20 bpm (Pediatric mode)
Doc. 007301 K-1