Breas Vivo 45 Quick Reference Manual
Hide thumbs Also See for Vivo 45:

Advertisement

Quick Links

Usage of Oxygen (Accessory)
O
must be turned off when the Vivo 45 is not in
2
operating mode!
It is recommended to monitor the oxygen concen-
trations.
Oxygen flow must not exceed 15 l/min.
1.
Mount the oxygen tube to
the connector.
2.
Connect the oxygen tube. A
click sound is heard when the
tube is properly attached.
Disconnect:
1.
Turn off the oxygen source.
2.
Press the release button to
disconnect the oxygen tube.
3.
Remove the tube.
Replacement of the Patient Air Inlet Filters
The filters are located in the filter
cassette.
Washable Filter (grey): Replace at
least once a year, wash at least
2
once a week.
Disposable Filter (white): Replace
at least every 4th week, or more
often in high pollution or pollen-
rich environments.
Symbols
S
D
YMBOL
Internal battery
Click-in battery level
Home Mode activated
Ramp function active
Humidifier and/or heated circuit connected
SpO2 sensor connected
FiO2 connected
EtCO2 connected
PtcCO2 connected
Effort belt connected
1
Multiple pages
Multiple content available
ESCRIPTION
ENGLISH
Vivo 45
Quick Reference Guide
Doc. 007610 En Y-1
2019-10-29