Temperature Comp. Lost; Compensation Lost Alarm); Humidity Comp. Lost; Lost Alarm) - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.4.17
Temperature Comp. Lost (Ambient Temperature Compensation Lost
Alarm)
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.18
Humidity Comp. Lost (Humidity Compensation Lost Alarm)
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
Doc. 007301 K-1
Description

Temperature Comp. Lost

An Ambient Temperature Compensation Lost alarm will be
given when the automatic ambient temperature compensa-
tion is out of order.
There is no communication with the air temperature sensor
or the value is out of range (less than -30°C (-22°F) or more
than 70°C (158°F).
Medium
The ventilator will continue treatment with the same settings.
The accuracy of the volume measurement may be impaired.
Ambient temperature inside valid range.
Description

Humidity Comp. Lost

An Humidity Compensation Lost alarm will be given when
the automatic humidity compensation is out of order.
50% relative humidity is used for temporary compensation.
If the ventilator is used at other humidities, delivered and
measured pressure and flow may deviate.
Medium
The ventilator will continue treatment with the same settings.
The accuracy of the volume measurement may be impaired.
Air humidity sensor values (RH and temperature) inside valid
range.
Alarms
145
User manual