High Minute Volume Alarm - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.3.5

High Minute Volume Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
96
Alarms
User manual
Description
High MV
Medium
A High Minute Volume alarm will be given when the
minute volume is above the set alarm limit for 15 seconds.
Disconnection of patient circuit.
Mismatch between pressure setting and alarm setting.
Leakage from the mask or other components of the patient
circuit.
Increased breath rate.
The minute volume goes below the alarm limit.
The Vivo 2 will continue treatment according to the cur-
rent settings.
1 l. to 40 l.
Off
0.5 l.
Doc. 007229 D-3