WARNING!
Do not use a humidifier between the oxygen source and the ventilator, in order to
humidify the oxygen flow.
If humidification is required, use an external humidifier after the patient air outlet.
WARNING!
Risk of Fire
Ventilate the room adequately. Do not smoke in a room where oxygen is being used.
Risk of Fire
Naked light bulbs and other sources of ignition must be kept a minimum of 2 meters
(6 feet) away from the oxygen cylinder or any part of the patient circuit.
Risk of Fire
Do not use aerosols or solvents close to the oxygen supply, even when the oxygen
supply is shut off.
WARNING!
Supplemental oxygen with a flow up to 15 l/min can be added by an oxygen source
with rotameter such as oxygen cylinder, central oxygen supply system.
Risk of faulty Treatment
At a fixed flow rate of supplemental oxygen flow, the inhaled oxygen concentration
will vary, depending on the pressure delivered, the patient's breathing pattern, the
patient interface and the leak rate.
To monitor the oxygen concentration, use the FiO
Supplemental oxygen flow and pressure must not exceed 15 l/min and 100 kPa.
32
Safety Information
User manual
sensor accessory.
2
Doc. 007301 K-1