High EtCO

Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
This alarm requires a connected EtCO
132
Alarms
User manual

Alarm

2
Description
High EtCO2
High
A High EtCO
EtCO
exceeds the alarm limit for 30 seconds.
2
Alarm limit is set too low.
Breath Rate too low.
Delivered Tidal Volume too low.
Excessive dead space between patient and exhalation valve/
leakage port.
Exhalation valve / leakage port is occluded.
1 to 99 mmHg
Off
1 mmHg
The ventilator will continue treatment with the same settings.
alarm will be given when the measured
2
sensor
2