6.4.15
FiO
Disconnected (FiO
2
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
Reset
6.4.16
Ambient Pressure Compensation Lost Alarm
Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset
144
Alarms
User manual
Sensor Failure/Disconnection Alarm)
2
Description
FiO2 Disconnected
An FiO
Sensor Failure/Disconnection alarm will be given
2
when no signal from the FiO
2 seconds. Check the FiO
High
•
FiO
Sensor disconnected.
2
•
Communication with the FiO
The ventilator will continue treatment with the same settings.
Confirmation by user or reconnected/changed.
Description
Pressure Comp Lost
Medium
An Ambient Pressure Compensation Lost alarm will be
given when the automatic ambient pressure compensation
functionality is out of order.
The Vivo 45 LS will continue treatment according to the cur-
rent settings.
Normal atmospheric pressure at sea level will be used as
approximation for the temporary ambient pressure compen-
sation. If used at other altitude, delivered and measured pres-
sures may deviate.
Reset of ventilator.
sensor has been detected for
2
sensor.
2
sensor failed.
2
Doc. 007301 K-1