6.4.6
Lost Mains Power Alarm
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.7
Exhalation Valve Control Error Alarm
Property
Alarm text
Alarm condition
Priority
Possible cause
Reset
140
Alarms
User manual
Description
Lost Mains Power
A Mains Power Lost alarm will be given when the ventilator
switched from Mains to another power source due to Mains
power is lost.
Medium
The ventilator will continue treatment with the same settings.
An information message will be shown on the screen.
Confirmation by user or mains reconnected.
Description
Exhalation Valve Control Error
An Exhalation Valve Control Error alarm will be given
when the ventilator fails to control the internal /external
exhalation valve.
High
•
Exhalation valve occluded
•
Exhalation valve control tube disconnected
•
Internal function failure of the exhalation valve controls
The pilot pressure gets a normal value.
Doc. 007301 K-1