Low Etco; Alarm - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

6.3.24

Low EtCO

Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
This alarm requires a connected EtCO
Doc. 007301 K-1

Alarm

2
Description
Low EtCO2
Medium
A Low EtCO
EtCO
is below the alarm limit for 30 seconds.
2
Alarm limit is set too high.
Ventilator disconnection.
Excessive leakage in the Patient circuit/Interface.
Partial obstruction of the airways.
Breath Rate too high.
Delivered Tidal Volume too high.
Self triggering of the ventilator.
1 to 99 mmHg
Off
1 mmHg
The ventilator will continue treatment with the same settings.
alarm will be given when the measured
2
sensor
2
Alarms
133