6.3.28
PtcCO
Property
Alarm text
Priority
Alarm Condition
Possible cause
Ventilator action
This alarm requires a connected PtcCO
Doc. 007301 K-1
Alarm
2
Description
PtcCO2 Outside Limits
High
A PtcCO
limits.
Check the PtcCO
External PtcCO
Breath Rate needs adjustment
Delivered Tidal Volume needs adjustment.
Excessive dead space between patient and exhalation valve/
leakage port.
Leak port/valve occluded.
Ventilator disconnection.
Excessive leakage in the Patient circuit/Interface.
Partial obstruction of the airways.
Self triggering of the ventilator.
The ventilator will continue treatment with the same settings.