6.3.28
PtcCO
Property
Alarm text
Priority
Alarm Condition
Possible cause
Ventilator action
This alarm requires a connected PtcCO
Doc. 007301 K-1
Alarm
2
Description
PtcCO2 Outside Limits
High
A PtcCO
limits.
Check the PtcCO
•
External PtcCO
•
Breath Rate needs adjustment
•
Delivered Tidal Volume needs adjustment.
•
Excessive dead space between patient and exhalation valve/
leakage port.
•
Leak port/valve occluded.
•
Ventilator disconnection.
•
Excessive leakage in the Patient circuit/Interface.
•
Partial obstruction of the airways.
•
Self triggering of the ventilator.
The ventilator will continue treatment with the same settings.
alarm will be given when PtcCO
2
monitor.
2
monitor is outside its alarm limits.
2
sensor.
2
is outside alarm
2
Alarms
137
User manual