Simv Rate; Insp. Time (Inspiratory Time) - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

5.5.6

SIMV Rate

Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
5.5.7

Insp. Time (Inspiratory Time)

Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
Doc. 007301 K-1
Description
The SIMV Rate setting is used in the SIMV ventilation modes, for
defining the minimum frequency of mandatory, ventilator-controlled
breaths. The mandatory breaths can be either triggered by an inspira-
tory effort from the patient, or ventilator-initiated. The SIMV Rate
setting determines the SIMV cycle time.
The combination of the SIMV Rate and Inspiratory Time setting is
limited by the I:E ratio 2:1.
4 bpm (Adult), 6 bpm (Pediatric)
40 bpm (Adult), 60 bpm (Pediatric)
1 bpm
12 bpm (Adult mode)
20 bpm (Pediatric mode)
Description
The Inspiratory Time setting defines the length of each inspiration
from start of inspiration to cycling off to expiration.
The combination of the Inspiratory Time and Breath Rate or SIMV
Rate settings is limited by the I:E ratio 2:1.
0.3 s
5 s (Adult), 2 s (Pediatric)
0.1 s
1.5 s(Adult mode)
1 s (Pediatric mode)