Immunity Test
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast tran-
sient/burst
IEC 61000-4-4
Surge IEC 61000-
4-5
Power frequency
(50/60 Hz) mag-
netic field
IEC 61000-4-8
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
U
is the mains voltage prior to application of the test level.
T
WARNING!
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the ventilator, including cables specified. Otherwise, degradation of the per-
formance of this equipment could result.
174
Technical Specifications
User manual
Compliance Level
±8 kV contact±15 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line to line
30 A/m
0% U
, 0.5 cycle (multiple
T
phase analysis);
0% U
, 1 cycle;
T
70% U
, 25/30 cycles (50/
T
60 Hz);
0% U
, 250/300 cycles (50/
T
60 Hz);
Electromagnetic Environ-
ment - Guidance
The relative humidity should
be at least 5 %.
Mains power quality should
be that of a typical commer-
cial, hospital and residential
environment.
Mains power quality should
be that of a typical commer-
cial, hospital and residential
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial, hospital and residential
environment.
Vivo 45 LS runs on internal
battery during voltage dips,
short interruptions and volt-
age variations on power sup-
ply input lines.
Doc. 007301 K-1