6.4.19

LED Failure Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.20

Low Alarm Battery Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.21

Alarm Battery Error Alarm

Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
146
Alarms
User manual
Description
LED Failure
A LED Failure alarm will be given when one or more LED
indicators on the front panel are broken.
Medium
The ventilator will continue treatment with the same settings.
Power-on reset of ventilator (or repair).
Description
Low Alarm Battery
An alarm for Low Alarm Battery will be given if the alarm bat-
tery is not charged enough to have power for a Power Fail
alarm for at least 2 minutes.
Medium
The ventilator will continue treatment with the same settings
and start charging the alarm batteries.
When alarm energy storage level is sufficient to give an alarm
for at least 2 minutes.
Description
Alarm Battery Error
Unable to communicate with super capacitor and read super
capacitor status.
Medium
The ventilator will continue treatment with the same settings.
When triggering condition is removed.
Doc. 007301 K-1