The ventilator always starts in the mode and with the settings that were active when
it was switched off.
4.7
Performing the Pre-use Test
The pre-use test is used for detecting the type and characteristics of the patient circuit
that is connected to the Vivo 45 LS. The resistance and compliance of the patient circuit
are measured and calculated. This will be used to compensate for pressure drop in the
patient circuit and the compliance of the patient circuit.
The patient shall not be connected during the pre-use test.
NOTE
If the pre-use test has not been performed, the Vivo 45 LS will operate with default
patient circuit compensation.
Starting the Pre-use Test Manually
1 On the Others menu, Select Pre-use Test and then Start Pre-use Test.
2 Follow the instructions on the display.
Activating the Pre-use Test Prompt
1 On the Others menu, Select Pre-use Test.
2 Set Pre-use Test to On.
52
Preparing the Vivo 45 LS for Use
User manual
Doc. 007301 K-1