Tidal Volume

ITEM
Definition
Setting min
Setting max
Setting
resolution
Doc. 007301 K-1
Description
The Target Volume setting defines the tidal volume that the ventilator
will aim for while ventilating the patient in a pressure mode. To aim
for the preset volume, the ventilator will adapt the Inspiratory Pres-
sure between two adjustable pressure limits: Min Pressure and Max
Pressure.
When Target Volume is active, the mode field on the ventilator display
will indicate "(TgV)".
Off
50 ml (Pediatric mode)
300 ml (Adult mode)
2000 ml (Adult mode),
400 ml (Pediatric mode)
10 below 500 ml,50 above 500 ml
Off
DESCRIPTION
The Tidal Volume setting defines the volume that will be delivered by
the Vivo 45 LS for each breath when using volume control modes.
In VCV-SIMV mode, this setting is applicable to the mandatory
breaths that are controlled by the ventilator.
300 ml (Adult mode)
50 ml (Pediatric mode)
2000 ml (Adult mode)
400 ml (Pediatric mode)
10 ml below 500 ml
50 ml above 500 ml