Breas Vivo 45 LS User Manual page 164

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

Ventilation modes
PSV – Pressure Support Ventilation
PSV(TgV) – Pressure Support Ventilation with Target Volume
PCV – Pressure Controlled Ventilation
PCV(TgV) – Pressure Controlled Ventilation with Target Volume
PCV(A) – Assisted Pressure Controlled Ventilation
PCV(A+TgV) – Assisted Pressure Controlled Ventilation with Target Volume
PCV-SIMV – Pressure Controlled Ventilation with Synchronized Intermittent Man-
datory Ventilation
PCV-MPV – Pressure Controlled Ventilation with MouthPiece Ventilation
VCV – Volume Controlled Ventilation
VCV(A) – Assisted Volume Controlled Ventilation
VCV-SIMV – Volume Controlled Ventilation with Synchronized Intermittent Man-
datory Ventilation
VCV-MPV – Volume Controlled Ventilation with MouthPiece Ventilation
CPAP – Continuous Positive Airway Pressure
Device modes
Clinical
Home
Patient modes
Adult
Pediatric
Insp. Pressure
Range/Performance: 4 to 50 cmH
tric mode.)
Tolerance: ±0.5 cmH
Resolution: 0.5 below 10 cmH
164
Technical Specifications
User manual
O. (Limited to 30 cmH2O above PEEP in Pedia-
2
O or ±5%, whichever is greatest.
2
O, 1.0 above 10 cmH
2
O
2
Doc. 007301 K-1