Regular Maintenance; Service And Repair; Storage - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

3 If a low resistance bacterial filter is used, it shall be replaced. To avoid cross-
contamination when no bacterial filter has been used, a validated ozone-disinfection
process may be used, see the section on disinfecting the main unit internally.
4 Use a new patient circuit when the ventilator is used by a new patient.
7.4

Regular Maintenance

Regular maintenance inspections and checks shall be carried out at least every 24
months, according to the ventilator Service Manual
WARNING!
Do not use the device and contact your responsible care provider for an inspection
of the device in the event of:
Unexpected patient symptoms during treatment.
Unexplainable or sudden pressure, performance or sound changes during
operation.
Suspected damage to the device, including the occurrence of Internal Functional
Failure alarms.
Suspected damage to the click-in battery, including evidence of battery cell leakage.
7.5

Service and Repair

The service and repair of the ventilator must only be carried out by authorised service
personnel in accordance with Breas service instructions. Service inspections must always
be carried out following any repairs to the device.
Authorised service workshops can order the ventilator Service Manual that contains
all technical documentation required for the maintenance and service of the
ventilator.
7.6

Storage

Store the ventilator in a dark room, where the temperature range is within -20 to +60°C
(-4 to +140°F).
160
Cleaning and Maintenance
User manual
Doc. 007301 K-1