5.3.18

Rise Time

Rise Time displays the duration of the pressure or volume increase, measured from the
start of inspiration until the set pressure or volume has been reached.
5.3.19

EtCO

2
EtCO
displays the end-tidal carbon dioxide, measured on the last portion of the
2
exhaled volume that is passing through the EtCO
5.3.20

InspCO

InspCO2 displays the inspired carbon dioxide.
5.3.21

PtcCO

2
PtcCO2 displays transcutaneous CO2 pressure from an external PtcCO2 monitor, if
connected by the PtcCO2 cable accessory.
5.4

Modes in the Vivo 45 LS

In the Modes section of the Vivo 45 LS display, the operator selects the ventilation
mode, breath mode, patient mode and device mode for the treatment.
5.4.1

Standby and Operating Mode

Standby mode is defined as the state of the Vivo 45 LS when it is powered on but with-
out having treatment or pre-use test started.
Operating mode is defined as the state of the Vivo 45 LS when the blower is operating
and producing an air flow.
Switch between operating and standby mode by starting/stopping the ventilator (see 5.1
Switch the Vivo 45 LS On and Off, page 54).
Some operations (such as setting time and date) are only available in standby mode.
5.4.2

Device Mode

The two device modes of the ventilator are used for controlling the user access to the
ventilator settings.
Clinical mode allows full access to the ventilator treatment parameters for health care
professionals.
Home mode is used to limit the access to the ventilator's settings for patients and lay
persons.
If the Home Adjust function is activated, the patient is given the possibility to change a
selection of patient parameters that is determined by the responsible care provider. The
limited settings for Home Mode are defined in the Setup section
70
How to Use the Vivo 45 LS
User manual
2
sensor.
2
Doc. 007301 K-1