6.3.1

High Pressure Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Setting Display
112
Alarms
User manual
Description
High Pressure
High
A High Pressure alarm will be given when the patient pres-
sure reaches the set High Pressure alarm limit for three con-
secutive breaths.
It will also be given if pressure exceeds 76 cmH
Mismatch between pressure setting and alarm setting.
Coughing during inspiration.
Changes in airway resistance and or compliance.
A full breath is performed with maximum pressure below
the alarm limit.
The Vivo 45 LS will continue treatment according to the
current settings. The actual breath is however terminated if
the High Pressure alarm limit is reached.
5 cmH
O to 70 cmH
2
Note that the High pressure alarm cannot be set lower than
the value set for the Low pressure alarm.
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
The High Pressure alarm setting is displayed by a red line in
the pressure bar graph.
O
2
O
2
O
2
O.
2
Doc. 007301 K-1