6.3.5
High Vt
Property
Alarm text
Priority
Alarm Condition
Possible cause
Reset criteria
Setting range
Setting resolution
Ventilator action
116
Alarms
User manual
(High Inspired Tidal Volume Alarm)
i
Description
High Vti
Medium
A High Inspired Tidal Volume alarm will be given when the
monitored Inspired Tidal Volume exceeds the set limit for
the High Inspired Tidal Volume alarm for 15 seconds.
•
Mismatch between Inspired Tidal Volume and alarm setting.
•
Pressure settings causing the Inspired Tidal Volume to exceed
the set alarm level.
•
Leakage from the mask or other components of the patient
circuit.
•
Mismatch between selected and used patient circuit.
When inspired tidal volume is below set alarm limit
•
Adult mode: 150 ml to 2500 ml
•
Pediatric mode: 30 ml to 600 ml
•
Off
10 below 600 ml, 100 above 600 ml
The ventilator will continue treatment with the same settings.
Doc. 007301 K-1