Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF IEC
61000-4-3
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Doc. 007301 K-1
Compliance Level
3 V
(150 kHz to 80 MHz)
rms
6 V
(inside ISM/ASR
rms
bands)
10 V/m 80 MHz to 2.5 GHz
Electromagnetic Environ-
ment - Guidance
d=0.35*√P m at 150 kHz to
80 MHz
d= 1.2*√P m at 80 MHz to
800 MHz
d= 2.3*√P m at 800 MHz to
2.5 GHz
Equation description: P is
the maximum output power
rating of the transmitter in
watts (W) according to the
transmitter manufacturer and,
except for portable RF com-
munications equipment, d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
a
survey
, should be less than
the compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with this symbol:
Technical Specifications
User manual
175