6.4.30

Bypass Loose Alarm

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset action
6.4.31

Alarm Speaker Fail

Property
Alarm text
Priority
Alarm condition
Ventilator action
Reset
6.5

Alarm Test

6.5.1

Alarm Signal Test

When starting treatment, an automatic alarm signal test is performed. Check that the
test is performed successfully, this is indicated by:
A short beep indicating functional audio signaling.
The alarm LED first lights yellow, then red, indicating functional visual signaling.
The audio pause LED lights yellow.
In about a second, both LEDs are turned off.
If the test fails, do not use the Vivo 45 LS. Contact your supplier of the Vivo 45 LS for
a technical check.
150
Alarms
User manual
Description
Bypass Loose
Medium
The Bypass Loose alarm shall be given when the air bypass
latch is stuck in the down position for 5 secs.
The ventilator will continue treatment with the same settings.
Reinsert the air bypass unit and make sure the latch closes.
Description
Alarm Speaker Fail
High
Failure detected to one of the alarm speakers.
The ventilator will continue treatment.
Push OK to clear the alarm. If the fault persists, contact your
service provider.