Backup Insp. Time (Backup Inspiratory Time); Sigh Rate - Breas Vivo 45 LS User Manual

Hide thumbs Also See for Vivo 45 LS:
Table of Contents

Advertisement

5.5.8

Backup Insp. Time (Backup Inspiratory Time)

Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
5.5.9

Sigh Rate

Item
Definition
Setting min
Setting max
Setting
resolution
Default
value
NOTE
In pressure modes (during the sigh breath), the high pressure alarm will automatically
be set 10 cmH
82
How to Use the Vivo 45 LS
User manual
Description
The Backup Inspiratory Time setting defines the length of each inspi-
ration delivered during ventilator-triggered backup ventilation, initi-
ated by the set Backup Rate.
The combination of the Backup Inspiratory Time and Backup Rate
setting is limited by the I:E ratio 2:1 (1:99 to 1:1 in MPV).
0.3 s
5 s (Adult), 2 s (Pediatric)
0.1 s
1.5 s(Adult mode)
1 s (Pediatric mode)
Description
The Sigh rate sets the frequency of which breaths with an increased
pressure or volume are delivered to the patient.
If the High Pressure alarm or the High Tidal Volume alarm is given,
the Sigh function will be disabled as long as the alarm condition
remains.
Off, every 10 breaths .
Every 250 breaths.
10 breaths
Off
O above set sigh pressure (max 70 cmH
2
O).
2
Doc. 007301 K-1

Advertisement

Table of Contents
loading

Table of Contents