Doc. 007301 K-1
Accuracy Error Alarm
Description
CO2 Accuracy Error
A CO
error in the CO
High
The ventilator will continue treatment with the same settings.
OK message from sensor or sensor disconnected by user.

Adapter Alarm

2
Description
Check CO2 Adapter
A Check CO
adapter is not attached correctly to the CO
replace the airway adapter.
High
The ventilator will continue treatment with the same settings.
OK message from sensor or sensor disconnected by user.
Sensor Error Alarm
Description
CO2 Sensor Error
A CO
CO
Replace the CO
formed in this condition.
High
The ventilator will continue treatment with the same settings.
OK message from sensor or sensor disconnected by user.
Accuracy Error alarm will be given when an accuracy
2
measurement has occurred.
2
Adapter alarm will be given when the airway
2
Sensor Error alarm will be given when an error in the
2
sensor has occurred.
2
sensor. CO
2
sensor. Check/
2
monitoring cannot be per-
2
User manual
Alarms
143